Osmometer

Special Features and Advantages

- Menu led operation on a wide illuminated LCD graphic display with touch operation (view video).
- Selectable languages. (German, English, French, Spanish, Portuguese, Swedish, Italian).
- Code protection (view Video) with Log-In and user administration for GLP-Assistance.
- Instrument-ID free configurable for identification to GLP-Assistance.
- Variable 2nd and 3rd calibration points.
- Single use plastic tubes.
- Wide input voltage range 100-230VAC suitable for all countries
- Reduced operating costs as no extra water supply is necessary and power consumption is low.


Operating of the Instrument

- Switch on instrument, wait for ready display (approx. 3 minutes). View Video
- Place 100 µl sample in the sample tube (± 10 %).
- Place the sample tube onto the measuring head.
- Lower the measuring head. The sample tube is dipped into the cooling slot.
- Enter sample number using the touch display. These entries can also be made after measurement.
- Superercooling is reached after 1 minute and the instrument gives an audible tone. The user dips the needle with ice crystals shortly into the sample and starts the freezing manually.
- As freezing point is reached, the display shows the measurement value in mosm / kg and the instrument gives an audible tone.
The measurement value is stored in the memory.
- Slide measuring head from cooling slot and remove the sample tube.
- Wipe the thermistor with a soft tissue.